JOURNAL ARTICLE

Miniature tri-axial force sensor for feedback in minimally invasive surgery

Abstract

A miniature tri-axial force sensor has been developed that complies with the requirements of integration into biomedical and robotic devices due to its size and accuracy. The sensor is a machined Titanium alloy body of the overall dimensions 8.6 × Ø3.4 mm with semi-conductor strain gauges embedded on it. The sensing beam's size is 1 × 1 × 4.4 mm. We present here the design, calibration method and performance evaluation of our sensor. Properly calibrated, it can provide absolute resolution of 5 mN at 2 N full scale. The accuracy of the estimated angle of incidence is ±1°. The measured magnitude of the force is within the ±4% range of the actual one. At maximal load the accuracy of the shear angle is ±1°.

Keywords:
Strain gauge Calibration Materials science Shear force Range (aeronautics) Conductor Beam (structure) Accuracy and precision Acoustics Optics Computer science Biomedical engineering Physics Engineering Composite material

Metrics

27
Cited By
1.53
FWCI (Field Weighted Citation Impact)
13
Refs
0.86
Citation Normalized Percentile
Is in top 1%
Is in top 10%

Citation History