Ning XueJeong‐Bong LeeSteven FolandSung Pil Chang
Telemetric sensing is a promising method to accomplish non-contact and continuous disease diagnoses. Implantable intraocular pressure (IOP) sensor using wireless telemetric technology is highly desirable. This paper presents a fully integrated SU-8-based passive wireless inductively-coupled pressure sensor, especially designed for IOP sensor. The sensor was microfabricated using biocompatible materials - gold and SU-8 - to form the LC parallel circuit and the pressure sensitive diaphragm. Multi-SU-8-layers were stacked to completely isolate the working circuit from the biological tissue medium. The characterization result of IOP sensor shows that it has relatively high responsivity (683 KHz/mmHg) in the pressure range of 0 – 60 mmHg and has pressure resolution much lower than 1 mmHg in both air and saline medium.
Ning XueSung Pil ChangJeong‐Bong Lee
Chengran WangZhenyuan LinChen‐Hsin SunVictor KohLinbo LiuXudong Chen
Po‐Jui ChenSaloomeh SaatiRohit VarmaMark S. HumayunYu‐Chong Tai
Po‐Jui ChenDamien C. RodgerSaloomeh SaatiMark S. HumayunYu‐Chong Tai