Ning XueSung Pil ChangJeong‐Bong Lee
Telemetric sensing has received a great deal of attention in noncontact human disease detection. To take advantage of this method, this paper presents a SU-8-based compact (1.52 mm × 3.23 mm × 0.2 mm) passive wireless pressure sensor, especially designed to measure the intraocular pressure. The sensor was microfabricated using biocompatible materials such as gold and SU-8 [1] to form the LC parallel circuit and the pressure sensitive diaphragm. The surface of the sensor is fully covered by SU-8, which isolates the working circuit from the biological tissue medium. The pressure signal can be detected by an external readout coil with up to 6 mm distance from the implanted sensor. The pressure sensitivity of the sensor was characterized in both air and saline environment. The microfabricated sensor has high sensitivity (>7,000 ppm/mmHg).
Po‐Jui ChenDamien C. RodgerSaloomeh SaatiMark S. HumayunYu‐Chong Tai
Ning XueSung‐Pil ChangJeong‐Bong Lee
Ning XueJeong‐Bong LeeSteven FolandSung Pil Chang
Po‐Jui ChenSaloomeh SaatiRohit VarmaMark S. HumayunYu‐Chong Tai
Edgar A. UnigarroDiego A. SanzAlejandro ArciniegasFernando RamírezFredy Segura-Quijano