Abstract

We report on the development of combined ultrasound and photoacoustic imaging system using Lead Zirconate Titanate (PZT) based flexible bulk transducer arrays. The transducer was made with a new and simple technique and electronically interfaced with a research ultrasound system to obtain ultrasound image of custom targets. We also report for the first time, A-line photoacoustic signal from the flexible transducer array thereby laying the foundation of body conformal ultrasound co-registered photoacoustic imaging system.

Keywords:
Transducer Photoacoustic imaging in biomedicine Ultrasound Lead zirconate titanate Materials science Photoacoustic Doppler effect Ultrasonic sensor Photoacoustic effect Acoustics SIGNAL (programming language) Conformal map Ultrasound imaging Biomedical engineering Computer science Optics Optoelectronics Medicine Physics Ferroelectricity Dielectric

Metrics

19
Cited By
1.41
FWCI (Field Weighted Citation Impact)
27
Refs
0.78
Citation Normalized Percentile
Is in top 1%
Is in top 10%

Citation History

Topics

Photoacoustic and Ultrasonic Imaging
Physical Sciences →  Engineering →  Biomedical Engineering