Shigeru ToyamaYukinao ChisuwaYoshihito Ikariyama
An amperometric glucose sensor based on carbon paste electrode containing glucose oxidase, nickelocene, and platinum particles was fabricated and characterized. The sensor was functional at a less extreme potential (-200 mV vs. Ag/AgCl) than those of amperometric mediator free sensors. The sensor worked not only in buffer solution but in serum. The output current was unaffected by ascorbate, whose concentration was about two orders of magnitude above the standard ascorbate concentration in human blood. Although the sensor response decreased gradually when native-enzyme was coblended, the sensor in which the enzyme was modified with methoxy-polyethylene-glycol-epoxy was much stabler due to the suppression of enzyme leakage than that with native-enzyme coblended.
Mitsuyasu KawakamiNoboru UriudaHidekazu KoyaShinichiro Gondo
Tokuji IkedaHirokazu HamadaKojiro MIKIMitsugi Senda
Plamen AtanasovA. KaishevaS. GamburzevI. IlievSergei Bobrin
Yang LiuHong TengHaoqing HouTianyan You