The use of a noninvasive hemodynamic monitor to estimate arterial stiffness, by measurement of pulse transit time (PTT), was demonstrated in microgravity. The monitor's utility for space applications was shown by establishing the correlation between ground-based and microgravity-based measurements. The system consists of a scale-based ballistocardiogram (BCG) and a toe-mounted photoplethysmogram (PPG). PTT was measured from the BCG I-wave to the intersecting tangents of the first trough and maximum first derivative of the PPG waveforms of each subject. The system was tested on a recent series of parabolic flights in which the PTT of nine subjects was measured on the ground and in microgravity. An average of 60.2 ms PTT increase from ground to microgravity environments was shown, and was consistent across all test subjects (standard deviation = 32.9 ms). This increase in PTT could be explained by a number of factors associated with microgravity and reported in previous research, including elimination of hydrostatic pressure, reduction of intrathoracic pressure, and reduction of mean arterial pressure induced by vasodilation.

Keywords:
Photoplethysmogram Arterial stiffness Biomedical engineering Pulse Wave Analysis Pulse wave Blood pressure Medicine Physics Computer science Internal medicine Optics

Metrics

6
Cited By
0.46
FWCI (Field Weighted Citation Impact)
16
Refs
0.70
Citation Normalized Percentile
Is in top 1%
Is in top 10%

Citation History

Topics

Cardiovascular Health and Disease Prevention
Health Sciences →  Medicine →  Cardiology and Cardiovascular Medicine
Non-Invasive Vital Sign Monitoring
Physical Sciences →  Engineering →  Biomedical Engineering
Spaceflight effects on biology
Health Sciences →  Medicine →  Physiology
© 2026 ScienceGate Book Chapters — All rights reserved.