Adriana M. PimentaM. Renata S. SoutoRita I. L. CatarinoM. Fernanda C. LealJosé L. F. C. Lima
Abstract An ofloxacin‐selective electrode was developed for application in pharmaceutical preparations, serum and urine analysis. The electrode is based on tetrakis[3,5‐bis(trifluoromethyl)phenyl]borate as ion exchanger in a poly(vinyl chloride) membrane. The sensor displayed a Nernstian response for ofloxacin over a wide concentration range (3×10 −6 –1×10 −2 mol L −1 ) with a slope of 55±1 mV per decade of concentration and a practical detection limit of 1×10 −6 mol L −1 . The electrode was successfully applied to the determination of ofloxacin in samples with no pretreatment steps, adequate accuracy and precision and recoveries within the intervals of 94–107 %.