JOURNAL ARTICLE

Fabrication and Characterization of Planar Screen-Printed Ag/AgCl Reference Electrode for Disposable Sensor Strip

Koutarou IdegamiMiyuki ChikaeNaoki NagataniEiichi TamiyaYuzuru Takamura

Year: 2010 Journal:   Japanese Journal of Applied Physics Vol: 49 (9R)Pages: 097003-097003   Publisher: Institute of Physics

Abstract

An accurate disposable planar Ag/AgCl reference electrode with an internal electrolyte was successfully fabricated by the screen-printing process. The internal electrolyte layer was also printed by using an electrolyte paste of sodium alginate containing KCl. The potential stability of the electrode was investigated at different operation times and Cl - concentrations in test solutions. Results show that the electrode has long-term potential stability (approximately 60 min), and that its performance does not depend on the Cl - concentration. This electrode can be used to provide various promising applications in sensing techniques based on disposable strips for sensing purposes.

Keywords:
Electrolyte Electrode Reference electrode Fabrication Materials science Planar Screen printing Characterization (materials science) Working electrode Optoelectronics Analytical Chemistry (journal) Nanotechnology Chemistry Composite material Chromatography Computer science

Metrics

22
Cited By
0.85
FWCI (Field Weighted Citation Impact)
25
Refs
0.74
Citation Normalized Percentile
Is in top 1%
Is in top 10%

Citation History

Topics

Analytical Chemistry and Sensors
Physical Sciences →  Chemical Engineering →  Bioengineering
Gas Sensing Nanomaterials and Sensors
Physical Sciences →  Engineering →  Electrical and Electronic Engineering
Biosensors and Analytical Detection
Physical Sciences →  Engineering →  Biomedical Engineering
© 2026 ScienceGate Book Chapters — All rights reserved.