Abstract

Label-free nanosensors can detect disease markers to provide point-of-care diagnosis that is low-cost, rapid, specific and sensitive. However, detecting these biomarkers in physiological fluid samples is difficult because of ionic screening. Here, we overcome this limitation by using distinct components within the sensor to perform purification and detection. 1 A microfluidic purification chip captures multiple biomarkers simultaneously from blood samples and releases them, after washing, into purified buffer for sensing by a silicon nanoribbon detector. This two-stage approach isolates the detector from the complex environment of whole blood, and reduces its minimum required sensitivity by effectively pre-concentrating the biomarkers. We show specific and quantitative detection of two model cancer antigens from a 10 uL sample of whole blood in less than 20 minutes.

Keywords:
Whole blood Biomarker Microfluidics Detector Nanosensor Point of care Chromatography Computer science Materials science Chemistry Nanotechnology Biology Medicine Immunology Pathology