JOURNAL ARTICLE

Effect of ethanol/trisodium citrate lock on the mechanical properties of carbothane hemodialysis catheters

A. L. Loomis BellRajesh JayaramanLavern M. Vercaigne

Year: 2006 Journal:   Clinical Nephrology Vol: 65 (05)Pages: 342-348   Publisher: Dustri-Verlag

Abstract

The effect of the ethanol/TSC lock on the catheters is unlikely to prohibit clinical use. After 9 weeks of exposure to the solution, the catheter segments could still be stretched to 22 times their length and withstand 11.5 kg (113 N) of force. Clinically produced forces during dialysis are many times smaller than the force required to break the catheters examined in this study. Therefore, the ethanol/TSC lock shows promise as a new catheter locking solution for the treatment of catheter-related infections. Further clinical studies are required.

Keywords:
Trisodium citrate Medicine Hemodialysis Catheter Elongation Catheter Hemodialysis Ultimate tensile strength Ethanol Surgery Biomedical engineering Composite material Nuclear chemistry Materials science Chemistry Biochemistry

Metrics

38
Cited By
2.50
FWCI (Field Weighted Citation Impact)
0
Refs
0.88
Citation Normalized Percentile
Is in top 1%
Is in top 10%

Citation History

Topics

Central Venous Catheters and Hemodialysis
Health Sciences →  Health Professions →  Emergency Medical Services
Vascular Procedures and Complications
Health Sciences →  Medicine →  Pulmonary and Respiratory Medicine
Infective Endocarditis Diagnosis and Management
Health Sciences →  Medicine →  Epidemiology

Related Documents

JOURNAL ARTICLE

Ethanol/Trisodium citrate for hemodialysis catheter lock

A. L. Loomis BellXuehong GuFrank J. BurczynskiLavern M. Vercaigne

Journal:   Clinical Nephrology Year: 2004 Vol: 62 (11)Pages: 369-373
JOURNAL ARTICLE

Effect of ethanol/trisodium citrate lock on microorganisms causing hemodialysis catheterrelated

Teresa TaklaSheryl ZelenitskyLavern M. Vercaigne

Journal:   The Journal of Vascular Access Year: 2008 Vol: 8 (4)Pages: 262-267
JOURNAL ARTICLE

Carbothane and Silicone Hemodialysis Catheters Maintain Structural Integrity After 26 Weeks of 70% Ethanol Lock Exposure

Journal:   American Journal of Kidney Diseases Year: 2013 Vol: 61 (4)Pages: B58-B58
JOURNAL ARTICLE

Locking catheters with trisodium citrate reduces infection during hemodialysis

Rebecca Doherty

Journal:   Nature Clinical Practice Nephrology Year: 2005 Vol: 1 (2)Pages: 64-64