JOURNAL ARTICLE

Polyaniline Langmuir−Blodgett Film Based Cholesterol Biosensor

Abstract

Cholesterol oxidase (ChOx) has been covalently linked to Langmuir-Blodgett (LB) monolayers of polyaniline (PANI)-stearic acid (SA) prepared onto indium-tin-oxide (ITO) coated glass plates via glutaraldehyde (Glu) chemistry. These ChOx/Glu/PANI-SA LB film/ITO electrodes have been characterized by FT-IR, cyclic voltammetry, and scanning electron microscopy, respectively. The results of response measurements carried out on these bioelectrodes using linear sweep voltammetry (LSV) reveal linearity from 25 to 400 mg/dL of cholesterol concentration with sensitivity of 88.9 nA mg(-1) dL. The linear regression analysis of bioelectrode reveals standard deviation and correlation coefficient of 0.737 microA and 0.9988, respectively. The low value of the Michaelis-Menten constant of these bioelectrodes obtained as 1.21 mM for the immobilized enzyme indicates increased interaction between ChOx and cholesterol in the PANI-SA LB film.

Keywords:
Cholesterol oxidase Cyclic voltammetry Polyaniline Langmuir–Blodgett film Chemistry Glutaraldehyde Biosensor Monolayer Nuclear chemistry Analytical Chemistry (journal) Linear sweep voltammetry Electrochemistry Chromatography Electrode Organic chemistry Physical chemistry Polymer Enzyme

Metrics

103
Cited By
8.98
FWCI (Field Weighted Citation Impact)
22
Refs
0.98
Citation Normalized Percentile
Is in top 1%
Is in top 10%

Citation History

Topics

Electrochemical sensors and biosensors
Physical Sciences →  Engineering →  Electrical and Electronic Engineering
Conducting polymers and applications
Physical Sciences →  Materials Science →  Polymers and Plastics
Analytical Chemistry and Sensors
Physical Sciences →  Chemical Engineering →  Bioengineering
© 2026 ScienceGate Book Chapters — All rights reserved.