JOURNAL ARTICLE

Failure analysis of fractured dental zirconia implants

Abstract

Abstract Objectives: The purpose of the present study was the macroscopic and microscopic failure analysis of fractured zirconia dental implants. Methods: Thirteen fractured one‐piece zirconia implants (Z‐Look3) out of 170 inserted implants with an average in situ period of 36.75±5.34 months (range from 20 to 56 months, median 38 months) were prepared for macroscopic and microscopic (scanning electron microscopy [SEM]) failure analysis. These 170 implants were inserted in 79 patients. The patient histories were compared with fracture incidences to identify the reasons for the failure of the implants. Results: Twelve of these fractured implants had a diameter of 3.25 mm and one implant had a diameter of 4 mm. All fractured implants were located in the anterior side of the maxilla and mandibula. The patient with the fracture of the 4 mm diameter implant was adversely affected by strong bruxism. By failure analysis (SEM), it could be demonstrated that in all cases, mechanical overloading caused the fracture of the implants. Inhomogeneities and internal defects of the ceramic material could be excluded, but notches and scratches due to sandblasting of the surface led to local stress concentrations that led to the mentioned mechanical overloading by bending loads. Conclusions: The present study identified a fracture rate of nearly 10% within a follow‐up period of 36.75 months after prosthetic loading. Ninety‐two per cent of the fractured implants were so‐called diameter reduced implants (diameter 3.25 mm). These diameter reduced implants cannot be recommended for further clinical use. Improvement of the ceramic material and modification of the implant geometry has to be carried out to reduce the failure rate of small‐sized ceramic implants. Nevertheless, due to the lack of appropriate laboratory testing, only clinical studies will demonstrate clearly whether and how far the failure rate can be reduced. To cite this article: Gahlert M, Burtscher D, Grunert I, Kniha H, Steinhauser E. Failure analysis of fractured dental zirconia implants. Clin. Oral Impl. Res . 23 , 2012; 287–293 doi: 10.1111/j.1600‐0501.2011.02206.x

Keywords:
Implant Materials science Fracture (geology) Dentistry Cubic zirconia Scanning electron microscope Maxilla Ceramic Medicine Composite material Surgery

Metrics

96
Cited By
4.86
FWCI (Field Weighted Citation Impact)
33
Refs
0.94
Citation Normalized Percentile
Is in top 1%
Is in top 10%

Citation History

Topics

Dental Implant Techniques and Outcomes
Health Sciences →  Dentistry →  Oral Surgery
Dental materials and restorations
Health Sciences →  Dentistry →  Orthodontics
Bone Tissue Engineering Materials
Physical Sciences →  Engineering →  Biomedical Engineering

Related Documents

JOURNAL ARTICLE

Failure analysis of fractured dental implants

Rajesh BansalAmit Raj SharmaVakil Singh Singh

Journal:   Metallurgical and Materials Engineering Year: 2022 Vol: 28 (3)Pages: 557-565
JOURNAL ARTICLE

Identification of failure mechanisms in retrieved fractured dental implants

Keren Shemtov‐YonaD. Rittel

Journal:   Engineering Failure Analysis Year: 2014 Vol: 38 Pages: 58-65
BOOK-CHAPTER

Zirconia for Dental Implants

Saurabh Gupta

ASM International eBooks Year: 2022 Pages: 479-485
JOURNAL ARTICLE

Immunologically Neutral Dental Implants Ceramic (Zirconia) Implants

Alan Arturo Zarzar

Journal:   Journal of Dentistry and Oral Sciences Year: 2022